Section 7

Troubleshootiing and functional checks

7.1 Troubleshooting

Fault

Possible Cause

Action

No display

Check power supply

Check that correct 9V ac power supply
is connected and switched on.

Erratic display

Check power supply

Unit must be used with supplied 9V
acpower supply. Usage of other units will
cause the 3510 not to operate.

Drifting erratic readings

Electrode fault

Use BNC cap to test 3510 (see 7.2)
Replace electrode.

Cannot calibrate

Electrode Fault

Use BNC cap to test 3510 (see 7.2)
Replace electrode.

ERROR EO ...

Buffer problem

Use freshly prepared buffers.

ERROR SLOPE ....

Electrode problem

Use BNC cap to test 3510 (see 7.2)
Replace electrode.

Unable to recognise

Using correct buffer set

Is the buffer type correct? Use AUTO for

buffers

Jenway supplied buffers.
Replace buffers
Use BNC cap to test 3510 (see 7.2)
Replace electrode.

Temperature readings

Temp probe faulty

Check 3510 using section 7.2

fluctuating

Replace temperature probe.

Temperature readings

Temp probe faulty

Check 3510 using section 7.2

incorrect

Replace temperature probe.

Manual temp not set

Set meter to read °C and set
temperature against a calibrated
thermometer.

Will not print

Connection

3510 requires the RS232 cable to
connect the printer.

Paper out

The feed light on the printer will flash if
the unit requires paper.

Battery flat

Connect ac power supply.

If the above does not answer your query try the FAQ section on the

www.Jenway.com

Website.

351 050/REV B/09-06