Setting the Time Constant

When there is rapid change in the measurement data, the readings

may become difficult to read. In such case, the speed of updating

the readings can be reduced by changing the time constant setting.

Time Constant determines the time span of the moving average.

When a larger (longer) time constant is selected, the readings will
be rather stable, and when a smaller (shorter) time constant is

selected, the readings will be more responsive and sensitive to the

change.

The time constant can be selected from 1, 5, 10 or 20 seconds.

< Time Constant Setting Screen >
By pressing the

key in a

measurement mode, the display
switches to the time constant setting
screen.


EXAMPLE: when setting the time constant to 20 seconds:
Select the time constant “20.0” by pressing

keys, and

press the

key.

















Time Constant

Time cons. unit

*

Time constant can

be set only for Air
Velocity and
Volumetric Flow
measurements.

*

The time constant

selected from 1, 5,
10 and 20 seconds
will be initialized
once the power is
turned off, and will
return to the
default setting of 1
second.