Setting the measurement unit and baud rate, 5 setting the measurement unit and baud rate – KANOMAX 6036 Multi-function thermal anemometer User Manual User Manual
Page 22

21
§ 5
Setting the Measurement Unit and Baud Rate
List of Measurement Units:
●
Air velocity: m/s, FPM
●
Volumetric flow rate:
m
3
/h, ft
3
/min, m
3
/min
●
Air temperature: °C, °F
●
Length: mm, inch
●
Measurement Mode
Press the
key to enter the unit
setting mode.
●
Length Unit
The unit to be set will blink.
Use
keys to select the unit.
Press the
key to save the setting
and to proceed to the next unit setting.
●
Air Velocity Unit
Repeat the above procedure for each
unit.
●
Volumetric Flow Unit
●
Air Temperature Unit
●
Baud Rate
Serial communication
baud rate used in this
instrument is shown
on the right.
4800bps
9600bps
19200bps
38400bps
*
The saved
setting will be
kept even
when the
instrument is
turned off.
*
To exit
(cancel) the
unit/baud rate
setting, press
the
key.
Any setting
saved before.
<Unit Conversion Table>
Air Velocity: 1m/s=196FPM
Air Temp: T (
O
F)=1.8×T(
°C
)+32
Volumetric Flow: 1 m
3
/h=35.32 ft
3
/h
Length: 1inch=25.4 mm
■
Serial
communication is
only available on
MODEL6036