Specifications, 8 specifications – KANOMAX 6036 Multi-function thermal anemometer User Manual User Manual

Page 28

Advertising
background image

27

§ 8

Specifications

Product

ANEMOMASTER Standard

ANEMOMASTER Professional

Model

6035

6036

Measuring Object

Clean Air Flow

Measuring
Range
(Resolution)

Air Velocity

0.01 to 30.0 m/s

(0.00 to 9.99m/s: 0.01m/s, 10.0 to 30.0m/s: 0.1m/s)

Air
Temperature

-20.0 to 70.0 °C

(0.1

o

C)

Pressure*

2

-5.00 to 5.00 kPa (0,01 kPa)

Duct Size Range

0 to 2550 mm
0 to 255 inch

Accuracy

Air Velocity

±3% of reading or ±0.015 m/s (±3 FPM), whichever is greater

Air
Temperature

± 0.5 °C

Pressure*

2

±3% of reading or ±0.01 kPa

Response Time

Air Velocity

Approx.1sec. (at Air Velocity: 1m/s, 90% Response)

Air
Temperature

Approx.30sec.(at Air Velocity: 1m/s, 90% Response)

Pressure*

2

Approx. 1 sec.

Sampling Functions

- Hold the reading
- Statistical calculation (Average, Maximum, Minimum)
- Time constant setting (1, 5, 10, 20 sec.)
- Remaining battery level indicator (4 levels)
- Selection of sampling units (Air velocity: m/s; Flow rate: m3/h,

m3/min, ft3/min; Air Temperature: °C or

o

F)

- Duct shape setting: Rectangular or Circular / Duct size unit: mm or

inch)

Data Storage

Max. 1500 data records.

External Dimensions

Probe: approx.Φ6(Φ14)x 294 to 1060mm Cable length: approx.2000

mm Main body: approx. 88(W) x 188(L) x 51(H) mm

本 体 :約 88(幅)×188(長さ)×51(厚み)mm

Output

Analog output*

1

: DC 0 to 3V

(Only for air velocity output)

Digital output:USB (automatically

switched to RS-232C when

connected to printer: Baud rate:

4800, 9600. 19200, 38400bps)

Analog output*

1

: DC 0 to 3V (Only

for air velocity output)

Power Source

Alkaline AA Batteries×6 (Manganese,

Ni-Cd, and Ni-MH batteries can

be used as well), AC Adapter*

3

: AC 100 to 240V (50/60Hz)

ACアダプタ※2:AC100to240V (50/60Hz)

Battery Life

Approx.10 hours (when Air velocity: 5 m/s, Air temp: 20°C, and using
Alkaline batteries)

Operating Environment

Main Body: 5 to 40°C (41 to 104

o

F) with no visible condensation

Probe: -20 to 70°C (-4 to 158

o

F) with no visible condensation

Storage Temperature Range

-10 to 50°C (14 to 122

o

F) with no visible condensation

Weight

Approximately

500 g

(batteries included)

Standard Accessories

Carrying case×1, Operation manualx1, Alkaline AA Batteries×6

USB cable×1, Measurement
software CDROM(for Windows) x 1

Optional Accessories

AC Adapter, Printer, Printer cable, ,Hands-free Case

*1: Analog Output is available on MODEL6035-A0, 6035-C0, 6036-A0, and 6036-C0.
*2: Pressure is available on MODEL6035-B0, 6035-C0, 6036-B0, and 6036-C0
*3: Optional

Advertising
This manual is related to the following products: