Technical data – KROHNE OPTIFLEX 1300C EN User Manual
Page 137

TECHNICAL DATA
8
137
OPTIFLEX 1300 C
www.krohne.com
03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en
8.4 Guidelines for maximum operating pressure (CRN certification)
Make sure that the devices are used within their operating limits. CRN certification is necessary
for all devices that are installed on a pressure vessel and used in Canada.
Figure 8-5: Pressure / temperature de-rating (ASME B16.5), Ø8 mm single cable probe, in °C and barg
Figure 8-6: Pressure / temperature de-rating (ASME B16.5), Ø0.3¨ single cable probe, in °F and psig
1 p [barg]
2 T [°C]
3 p [psig]
4 T [°F]
5 Threaded connection, NPT (ASME B1.20.1).
6 Flange connection, Class 300 and Class 600. Threaded connection, NPT (ASME B1.20.1).
7 Flange connection, Class 150
-50
0
50
100
150
200
0
5
10
15
20
25
30
35
40
38
-60-50
0
50
100
150
200 212
250
300
350
392 400
0
50
100
150
200
250
300
350
400
450
500
550
600