Operation, 5 function description – KROHNE OPTIFLEX 1300C EN User Manual

Page 73

Advertising
background image

OPERATION

6

73

OPTIFLEX 1300 C

www.krohne.com

03/2014 - 4000172306 - HB OPTIFLEX 1300 R07 en

6.3.5 Function description

A. Quick Setup

Menu
No.

Step

Function

Function description

Selection list

Default

A.1 Setup Mode

A.1.1

A.1.1

A.1.1

A.1.1

Complete

Complete

Complete

Complete

This follows the steps given in

This follows the steps given in

This follows the steps given in

This follows the steps given in

the Application, Installation,

the Application, Installation,

the Application, Installation,

the Application, Installation,

Conversion and Output setup

Conversion and Output setup

Conversion and Output setup

Conversion and Output setup

modes.

modes.

modes.

modes.

A.1.2

A.1.2

A.1.2

A.1.2

Application

Application

Application

Application

Follow this procedure to

Follow this procedure to

Follow this procedure to

Follow this procedure to

describe the tank contents.

describe the tank contents.

describe the tank contents.

describe the tank contents.

1

Probe Type

Read only

2

Product Type:

Which application

do you have?

The application.

Liquid, Solid

Depends on

the probe

type

3

Air Gap Flag: Do

you have air or gas

above the product?

This tells the device whether or

not there is air or gas above the

product.

Air Gap, No Air Gap

1

Air Gap

4

Number Of

Products: How

many different

products do you

have in the tank?

The number of products in the

tank.

1, 2, >2

2

1

5

Top Product

Liquid:

Liquid:

Liquid:

Liquid: Alcohol, Base,

Hydrocarbon, Liquid Gas,

Mineral Oil, Solvent,

Aqueous Solution, Other

Solid:

Solid:

Solid:

Solid: Cereals, Minerals,

Plastic, Other

6

Application Type:

What do you want

to measure?

Level, Interface, Level +

Interface

3

Level

7

Are the products

mixed or unmixed?

If there are 2+ liquids in the

tank, is there an emulsion in

the tank. This tells the device

to look for the interface when

the two liquids have settled.

Mixed, Unmixed

Unmixed

8

Settings summary

9

Current changes

must be saved or

cancelled before

proceeding

Save, Cancel

4

A.1.3

A.1.3

A.1.3

A.1.3

Installation

Installation

Installation

Installation

Follow this procedure to

Follow this procedure to

Follow this procedure to

Follow this procedure to

describe the process

describe the process

describe the process

describe the process

connection.

connection.

connection.

connection.

1

Tank Height

The distance from the tank

connecting flange face/thread

stop down to the tank bottom.

min-max:

0.165…60 m /

0.54…196.86 ft

5

2

Installation Type

The type of process connection

on the tank for the device.

Socket, Stillwell (A),

Nozzle (B)

nozzle

3A

Stillwell Height

The height from the tank

connecting flange face/thread

stop to the bottom of the

stillwell.

min-max:

0…60 m / 0…196.86 ft

1.5 m /

4.92 ft

4A

Stillwell Diameter

The inner diameter of the

stillwell.

min-max:

20…1000 mm /

0.79…39.37 in

100 mm /

3.94 in

Advertising