The default value of the custody transfer protection code when leaving the factory is

↵→↑↵↑→↵→↑

This password may be changed using Menu 3.8.7.. However, this can only be done if custody
transfer protection is disabled first, as described above.

Begin from measuring mode.

If the two entered passwords are different the message "CODE WRONG" is displayed.

The message has to be acknowledged with the

↵

key and input must be repeated with Fct.

3.8.7.. Then the protection state 'active' or 'inactive' can be chosen with Fct.3.8.6..

NOTE:
If the input of the custody transfer password is wrong, a 9-character code is displayed. With this
character code the manufacturer can decode the password in cases where the password is
lost.

It is also possible to protect just the mass totaliser. Menu 3.8.5. ENABL.RESET determines
whether the operator can clear the mass total from the ACKNOWLEDGE/RESET MENU.

Begin from measuring mode.

Key

Display
line 1

line 2

→

Fct. (1).0

OPERATOR

2x

↑

Fct. (3).0

INSTALL.

→

Fct. 3.(1).0

BASE PARAM.

7x

↑

Fct. 3.(8).0

USER DATA

→

Fct. 3.8.(1)

LANGUAGE

6x

↑

Fct. 3.8.(7)

CODE 3

→

CodeE 3

- - - - - - - -

Enter new custody transfer
password two times.
Fct. 3.8.(7)

CODE 3

4x

↵

Key

Display
line 1

line 2

→

Fct. (1).0

OPERATOR

2x

↑

Fct. (3).0

INSTALL

→

7x

↑

Fct. 3.(8).0

USER DATA

→

4x

↑

Fct. 3.8.(5)

ENABL.RESET

→

(YES)

↑

(NO)

↵

Fct. 3.8.(5)

ENABL.RESET

4x

↵

+110.25

kg

Mass Total Display

↵

CodE 2

- -

↑→

RESET MASS

→

BLOCKED

Totaliser Reset is disabled

↵↵