KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 79

Advertising
background image

79

To test the frequency, connect a frequency meter to the pulse output terminals and proceed as
follows:

To test the pulse output, connect an external counter to the output terminals. When testing the
pulse output, the operator has the choice of the following pulse widths: 0.4 ms, 1.0 ms, 10.0
ms, 100.0 ms and 500 ms. The operator should choose the pulse width that best matches the
performance of his external pulse counter.

Connect a pulse counter to the impulse output and proceed as follows:

Fct. 2.(3).0

TEST P

→

SURE (NO)

↑

SURE (YES)

↵

Fct. 2.3.(1)

FREQUENCY

↑

Fct. 2.3.(2)

TEST PULSE

→

(0.4 mSec)

Use the

↑

to select the

desired pulse width

↑

(1.0 mSec)

↑

(10.0 mSec)

↑

(100.0 mSec)

Having selected the pulse
width zero the external
counter and then press

↵

↵

625

100.0 mSec

Key

Display
line 1

line 2

Fct. 2.(2)

TEST I

↑

Fct. 2.(3)

TEST P

→

SURE (NO)

↑

SURE (YES)

↵↵

Fct. 2.3.(1)

FREQUENCY

→

(LEVEL LOW)
0V on the output

↑

(LEVEL HIGH)
+V on the output

↑

1 Hz
A frequency meter
connected to the
output will show 1Hz.

↑

10 Hz

↑

100 Hz

↑

1000 Hz
After testing the

1000 Hz signal
connect a counter to
the output.

↵

Fct.2.3.(1)

FREQUENCY