BM 70M-EEx/ BM 70Mi-EEx

15

Assembly screws (M) for signal converter:

M

M

Purge connection
(optional)

Purge
connection

(optional)

Sensor connection
(optional)

Sensor connection
(optional)

H

H

O-Ring

O-Ring

Signal converter

Flange system

SA

Flange system

V96

9.2

Replacement of complete device

The requirements described in Sect. 9.1 also apply in this instance.
In addition, make sure that all process connections and the tank are non-pressurized.

Caution:

On pressurized tanks, do not remove the 4 screws (H) connecting the wave
guide window to the tank mounting flange (refer to diagram in Sec. 9.1).
DANGER TO LIFE!

In connection with environmentally critical products, carefully decontaminate the wetted parts of
the flange system after dismantling.

10 Maintenance

Service

Maintenance work of a safety-relevant nature within the meaning of explosion protection may only
be carried out by the manufacturer, his authorized representative or under the supervision of
authorized inspectors.