KROHNE BM 70 M User Manual

Page 60

Advertising
background image

60

BM 70 M

Fct. 2.1.2. DISPLAY

•

Select Function 2.1.2, as described in Sect. 8.2 and 8.3.

•

Press the

→

key; all segments in the display are switched on.

•

Press

↑

key; all segments are switched off.

•

This procedure (switching on/off) can be repeated any number of times with the

↑

key.

•

Press the

↵

key to terminate the test.

Fct. 2.1.3 STATUS

After selecting "Module DIS" or "Module MW" after pressing the

↵

key in each case an Identity

Number and an 8-digit binary number are displayed. If an undefined error occurs, please inform
KROHNE Service of these values. Terminate the test by pressing the

↵

key twice.

8.7.3 Test of current output

Fct. 2.2.1 VALUE I

•

Select Function 2.2.1, as described in Sect. 8.2 and 8.3.

•

Press

→

key to display the actual value of the current output (terminals 31 and 32) in "mA".

•

Terminate test by pressing the

↵

key.

Fct. 2.2.2 TEST I

•

A milliamperemeter should be connected to terminals 31 and 32 for this test, see Sect. 7.8

•

Select Function 2.2.2, as described in Sect. 8.2 and 8.3.

•

Press

→

key.

•

Safety inquiry: SURE NO

SURE YES

Select with

↑

key

•

After SURE YES, press the

↵

key for the 1st value in the following list to be present at the

current output.

•

Select current values with the

↑

key:

-

2 mA

-

4 mA

-

6 mA

-

8 mA

-

10 mA

-

12 mA

-

14 mA

-

16 mA

-

18 mA

-

20 mA

-

22 mA

•

The connected milliamperemeter indicates the selected current value.

•

For instruments with Ex-i current output: in the 2 mA setting the minimum current of 3.6 mA is
output.

•

Terminate the test by pressing the

↵

key, and the actual measured value is again present at

the current output.