8.7.4 Test of switching output

Fct. 2.3.1 TEST S

In this test, the switching contact (terminals 41 and 42) can be closed or opened.

Select Function 2.3.1 as described in Sect. 8.2 and 8.3.

Press the

key.

Safety inquiry: SURE NO

SURE YES

Select with

key

After SURE YES, press

key.

Displayed: OPEN = switching contact is open.

Press the

key.

Displayed: CLOSED = switching contact is closed.

Terminate the test by pressing the

key, and the switching output will resume its normal

operating condition.

8.7.5 Display of firmware version

Fct. 2.4.1 MASTER

This function shows the current main firmware version (e.g. V.4.04) in the configuration menu.

Fct. 2.4.2 DISPLAY

This function shows the current firmware version of the display part.

8.8

Pointers and error messages during measurement

8.8.1 Different kinds of messages

a) Pointers (operating conditions) during measurement provided by 6 markers in the 3rd (bottom)
line of the display, see Sect. 8.8.2.

b) Application errors, indicating an application error condition in which no valid measured values
are currently established. By changing the tank conditions (e.g. filling) this error can usually be
remedied, see "Error list" in Sect. 8.8.3.

c) Minor errors, indicating the failure of functions that usually do not affect measurement, see
"Error List" in Sect. 8.8.3.

d) Correctable errors, which can be remedied by operator action (manual) at the BM 70 M, or
possibly by KROHNE Service, see "Error List" in Sect. 8.8.3.

e) FATAL ERROR renders operation of the BM 70 M impossible; occurs only when the device is
switched on (powered) or after special test functions have been executed, when the self-test
identifies an error in the electronics of the BM 70 M, see "Error List" in Sect. 8.8.3.