Command #131

Write Meter Size

Changes of the converter meter size will force the updating of sensor characteristics (returned by Command #14) and
therefore might cause the automatic correction of the Primary Variable Upper Range Value (Menus 1.1.1, 3.1.1) and
PulseValue (Menu 3.4.3). Master is prompted about corrections carried out via warning responses and must reread the
updated data. For additional information refer to Command #35.

REQUEST DATA BYTES

#0

#1

#2

#3

METER

METER

SIZE

SIZE

MSB

LSB

RESPONSE DATA BYTES

#0

#1

#2

#3

METER

METER

SIZE

SIZE

MSB

LSB

Data Byte #0-#3

Meter size, IEEE 754, in meters. Allowed settings: 0.025 to 4 meters (0.98 to 157.48 inches).

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1-2

Undefined

3

Passed Parameter too Large

4

Passed Parameter too Small

5

Too Few Data Bytes Received

6

Local Device User

7

In Write Protect Mode

8-113

Undefined

114

Range Exceeded Max and Was Corrected

115

Range Surpassed Min and Was Corrected

116

Both Range and Pulse Value Corrected

117-127 Undefined