KROHNE UFC 500 HART User Manual

Page 25

Advertising
background image

HARTô Smart Communications Protocol

UFC500 Command Specification Rev: 2.0 Release Date: 9 Jun 2001

© 2001 KROHNE Mefltechnik D-47058 Duisburg

Page

25

8.10

Command #140

Read Current Output Parameters

Reads a set of parameters for device current output.

REQUEST DATA BYTES

NONE

RESPONSE DATA BYTES

#0

#1

#2

#3

FIXED

CURRENT

CURRENT

CURRENT

VAR

MAX

ZERO

HUNDRED

RANGE

PCT

PCT

INDIC

FLOW

FLOW

Data Byte #0

Fixed/variable range indicator:
0 - Variable range,
1 - Fixed range 0 - 20 mA,
2 - Fixed range 4 - 20 mA.

Data Byte #1

Maximal possible output current, in mA.

Data Byte #2

The value of current in mA, that corresponds to 0% flow (percents are related to Primary
Variable Upper Range Value).

Data Byte #3

The value of current in mA, that corresponds to 100% flow (percents are related to Primary
Variable Upper Range Value for the forward flow, or to the Reverse Range for the backward
flow).

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1-5

Undefined

6

Local Device User

7-127

Undefined