KROHNE UFC 500 HART User Manual

Page 31

Advertising
background image

HARTô Smart Communications Protocol

UFC500 Command Specification Rev: 2.0 Release Date: 9 Jun 2001

© 2001 KROHNE Mefltechnik D-47058 Duisburg

Page

31

8.15

Command #147

Write Flow Direction And Primary Head Constant

Writes direction of the forward flow and the primary head constant.

REQUEST DATA BYTES

#0

#1

#2

#3

#4

FLOW

PRIMARY

PRIMARY

DIR

HEAD

HEAD

VALUE

VALUE

MSB

LSB

RESPONSE DATA BYTES

#0

#1

#2

#3

#4

FLOW

PRIMARY

PRIMARY

DIR

HEAD

HEAD

VALUE

VALUE

MSB

LSB

Data Byte #0

defines direction of the forward flow:
0 - Positive (corresponds to м+о arrow on primary head),
1 - Negative.

Data Byte #1-#4

Primary head constant GK, IEEE 754. See primary head nameplate. Range: 0.5 -14.0.

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1

Undefined

2

Invalid Selection (of Flow Direction)

3

Passed Parameter too Large (Primary Constant)

4

Passed Parameter too Small (Primary Constant)

5

Too Few Data Bytes Received

6

Local Device User

7

In Write Protect Mode

8-127

Undefined