Martel Electronics PSC4010 User Manual

Page 21

Advertising
background image

19

5.3-4 Calibrator Status

Status registers, enable registers, and queues provide status information on the calibrator.
Each status register and queue has a summary bit in the Serial Poll Status Byte. Enable
registers generate summary bits in the Serial Poll Status Byte. The following is a list of
registers and queues along with their function.

Serial Poll Status Byte (STB)

The STB is sent when the calibrator responds to the *STB? command. Figure 24
demonstrates how it functions. Cleared when power is reset.

Service Request Enable Register (SRE)

Enables or disables the bits of the STB. Cleared when power is reset. Setting bits to 0
disables them in the STB. Setting the bits to 1 enables them. Bit assignments for the SRE and
the STB are shown below.

7

6

5

4

3

2

1

0

0

MSS

ESB

0

EAV

0

0

0

MSS

Master Summary Status. Set to 1 when ESB or EAV are 1 (enabled). Read using
the *STB? command.

ESB

Set to 1 when at least one bit in ESR is 1.

EAV

Error Available. An error has been entered into the error queue, and may be read
using the Fault? command.

Event Status Register (ESR)

A two-byte register, in which the lower bits represent conditions of the Calibrator. Cleared
when read and when power is reset.

Event Status Enable Register (ESE)

Enables and disables bits in the ESR. Setting a bit to 1 enables the corresponding bit in the
ESR, and setting it to 0 disables the corresponding bit. Cleared at power reset. Bit
assignments for the ESR and the ESE respectively are shown below.

15

14

13

12

11

10

9

8

0

0

0

0

0

0

0

0

Advertising