Specifications – Martel Electronics PSC4010 User Manual

Page 30

Advertising
background image

28

6. Specifications

One year at 23°C ± 5°C, unless specified otherwise. Outside of this range the stability of the
measurements is ± 0.003% of FS/°C.

Table 9: General Specifications

Operating Temperature

-10°C to 50°C

Storage Temperature

-20°C to 60°C

Power

4 X AA batteries; Alkaline or optional rechargeable

Battery Life

Approximately 20 hours

Serial Communications

ASCII; RS232 or USB depending on the communications cable used

CE - EMC

EN61326: 2006, ANNEX B Criterion A

Safety

CSA C22.2 No. 1010.1: 1992

Size (with holster)

8.5” x 4” x 2” (21.6cm x 10.2cm x 5.1cm)

Weight

Approximately 28oz. (0.8kg)

Table 10: DC Voltage Measurement/Source

Range

Accuracy (% of reading ± floor)

Read

0.000V - 20.000V

0.015% ± 2mV

Source

0.000V - 20.000V

0.015% ± 2mV

Maximum current output in voltage ranges is 3mA with an output impedance of <= 1

Ω.

Table 11: Millivolts Measurement/Source

Range

Accuracy (% of reading ± floor)

Read (mV)

0.000mV - 90.000mV

0.02% ± 10

μV

Source (mV)

0.000mV - 100.000mV

0.02% ± 10

μV

Maximum current output in voltage ranges is 1mA with an output impedance of <= 1

Ω

Table 12: DC mA Measurement/Source

Range

Accuracy (% of reading ± floor)

Read

0.000mA - 24.000mA

0.015% ± 2

μA

Source

0.000mA - 24.000mA

0.015% ± 2

μA

Maximum load on mA source is 1000

Ω. Voltage input range on simulate mode 5V - 30V.

For units configured to support 50mA the table below shows the specification.

Range

Accuracy (% of reading ± floor)

Read

0.000mA - 52.000mA

0.02% ± 4

μA

Source

0.000mA - 52.000mA

0.02% ± 4

μA

Maximum load on mA source is 400

Ω at 52mA. Voltage input range on simulate mode 5V - 60V.

Advertising