Martel Electronics PSC4010 User Manual

Page 28

Advertising
background image

26

LOCAL

Restores the calibrator to local operation if it was in remote mode. Also clears LOCKOUT if the
unit was in lockout mode.

MEAS

Sets the display to measure mode. The command is followed by any of the parameters
except for pulse and mA sim, which are source only. Enter DCI for mA, DCV for volts, DCMV
for millivolts, FREQUENCY for frequency, and PRESSURE for pressure. For example:

MEAS DCV sets the display mode to VOLTS IN

OUT

Sets the output of the calibrator. This command may be used to output mA, volts, frequency,
temperature, and ohms. Frequency output, which is set by the command FREQ_TYPE, is
either continuous or pulse. The calibrator is automatically set to source mode when OUT is
entered. A number and its unit must follow the command. See Table 7. (Parameter Units) for a
list of available units. For example:

OUT 10 MA sets the calibrator to mA OUT mode and sets the output to 10mA.

OUT?

Returns the output of the calibrator. Using the above example, OUT? Would return:

1.000000E-02, A

PULSE_CNT

Sets the number of pulses the calibrator will produce when it is triggered while in pulse mode.
For example;

PULSE_CNT 3000 will set the number of pulses to 3000.

PULSE_CNT?

Returns the number of pulses in the pulse train. Using the above example, the returned value
would be:

3000

REMOTE

Puts the calibrator in remote mode. From the remote mode the user can still use the keypad
to get back to local unless the command LOCKOUT was entered before REMOTE. Than the
keypad is totally locked out, and the user has to send the LOCAL command to get back to
local operation.

Advertising