Command

Description

PULSE_CNT?

Returns the number of pulses in the pulse train

REMOTE

Puts the calibrator in remote mode

SIM

Sets the output for mA simulation

SIM?

Returns the output of the mA simulation

TRIG

Starts and stops the pulse train in pulse mode

TRIG?

Returns TRIGGERED when a pulse train is on. Returns UNTRIGGERED when the pulse
train is off.

VAL?

Returns the measured values

V_RANGE

Sets voltage range

V_RANGE?

Returns voltage range

Table 7: Parameter units

Units

Meaning

MA

milliamps of current

MV

Voltage in millivolts

V

Voltage in volts

CPM

Frequency in cycles per minute

Hz

Frequency in Hertz

KHz

Frequency in kiloHertz

Table 8: Error codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a numeric entry

101

Too many significant digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

110

An unknown command was received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

115

Output is overloaded

116

Calibrator not in pulse train mode when TRIG was received

117

An invalid FREQ_TYPE was received