10. Specifications

All measurements apply at 23°C ± 5°C. unless specified otherwise.
Outside of this range the stability of the measurements is ±
0.005%of reading/°C.

Table 9: General Specifications

Operating Temperature

-10°C to 50°

Storage Temperature

-20°C to 70°C

Power

4 X AA batteries; Alkaline or optional rechargeable

Low battery warning

Yes

Serial Communications

Yes, ASCII

CE - EMC

EN50082-1: 1992 and EN55022: 1994 Class B

Safety

CSA C22.2 No. 1010.1: 1992

Table 10: DC Voltage Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read: Isolated(Upper Display)

0.000V- 30.000V

0.015% ± 2mV

Read: non-Isolated(Lower Display)

0.000V - 20.000V

0.015% ± 2mV

Source

0.000V - 20.000V

0.015% ± 2mV

Maximum current output in voltage ranges is 3mA with an output impedance of <= 1

W.

Table 11: DC mA Measurement/Source

Range

Accuracy(% of reading

± floor)

Read: Isolated(Upper Display)

0.000mA - 24.000mA

0.015% ± 2

mA

Read: non-Isolated(Lower Display)

0.000mA - 24.000mA

0.015% ± 2

mA

Source

0.000mA - 24.000mA

0.015% ± 2

mA

Maximum load on mA source is 1000

W. Voltage input range on simulate mode 5V - 30V.

Table 12: Frequency Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read

2.0CPM - 600.0CPM

0.05% ± 0.1CPM

1.0Hz - 1000.0Hz

0.05% ± 0.1Hz

1.00KHz - 10.00KHz

0.05% ± 0.01KHz

Source

2.0CPM - 600.0CPM

0.05%

1.0Hz - 1000.0Hz

0.05%

1.00KHz - 10.00KHz

0.125%

Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude
is adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight
negative offset of approximately -0.1V is present to assure zero crossing.