Martel Electronics DMC1400 User Manual

Page 29

Advertising
background image

Figure 18. Measuring Pressure Transmitter

Note: On high pressure modules engineering units normally
associated with low pressure ranges such as, inH

2

O, cmH

2

O, etc. are

not valid selections. Selecting one of these units with a high
pressure module attached will cause the display to read "----".

6. Using the Upper and the Lower Display for

Calibration and Testing

6.1 Testing an Input or Indicating Device

To test and calibrate actuators, recording, and indicating devices
using the source functions, follow these steps:

1.

Select the lower display and choose the correct primary
parameter.

2.

Switch to [OUT] in the input/output control.

3.

Connect the leads to the instrument and the calibrator as shown
in Figure 19.

Figure 19. Connections for Testing an Output Device

27