CPRT_COEFB?

Returns the number, which was entered for the first coefficient of the
polynomial used in the custom RTD. Using the example above,
CPRT_COEFB? Would return:

-5.801900E-07

CPRT_COEFC

This command is used for entering a custom RTD into the calibrator.
The numeric value entered after the command will be set as the first
coefficient of the polynomial used by the custom RTD.

For example:

CPRT_COEFC -5.8019E-12 enters -5.8019e-12 as coefficient A.

CPRT_COEFC?

Returns the number which was entered for the first coefficient of the
polynomial used in the custom RTD. Using the example above
CPRT_COEFC? Would return:

-5.801900E-12

CPRT_MIN_T

Sets the minimum temperature of the custom RTD range. The
temperature value must be entered with a degrees label, CEL for
Celsius and FAR for Fahrenheit.

For example:

CPRT_MIN_T -260 CEL enters -260°C as the minimum
temperature.

CPRT_MIN_T?

Returns the value entered for minimum temperature in the range for
a custom RTD. Note that the Calibrator always returns numbers in
scientific notation. The above example would return:

-2.600000E+02, CEL

CPRT_MAX_T

Sets the maximum temperature of the custom RTD range. The
temperature value must be entered with a degrees label, CEL for
Celsius and FAR for Fahrenheit.

For example:

CPRT_MAX_T 0.0 CEL enters 0.0°C as the maximum
temperature.