Martel Electronics DMC1400 User Manual

Page 85

Advertising
background image

The letter U indicates an uncalibrated downloaded tag, the remainder
have been calibrated. Tag numbers are displayed on multiple pages, six
tags to a page. Press the UP ARROW and DOWN ARROW keys to
scroll the tag selector highlight one tag at a time, scrolling from page to
page, cycling from top to bottom of the list and vice versa. Use the LEFT
ARROW and RIGHT ARROW keys to scroll the tag selector highlight one
page at a time, cycling from top to bottom of the list and vice versa.

Press the EXIT function key to return to the review selection menu with-
out selecting a tag.

When the desired tag number is highlighted, press the SELECT function
key to display the first page of data showing the tag number, calibration
date and time, and the first few identification fields.

Press the UP ARROW and DOWN ARROW keys to scroll through the
remaining pages of data, cycling from the last page to the first page and
vice versa. The second page contains the remaining identification fields
and the validation error tolerance. The third page contains the instrument
input setup. The fourth page contains the instrument output setup. The
fifth and subsequent pages display the test results. On these pages only,
use the LEFT ARROW and RIGHT ARROW keys to switch between the
As Found results and the As Left results.

Note that the input/output orientation of the data on these pages is with
respect to the instrument under test. This is in contrast to the orientation

83

Advertising