5 entering commands, Table 8: error codes – Martel Electronics DMC1400 User Manual

Page 41

Advertising
background image

InH2O4C

Pressure in inches of water at 4°C

InH2O20C

Pressure in inches of water at 20°C

CmH2O4C

Pressure in centimeters of water at 4°C

CmH2O20C

Pressure in centimeters of water at 20°C

Bar

Pressure in bars

Mbar

Pressure in millibars

KPal

Pressure in kiloPascals

InHg

Pressure in inches of mercury at 0°C

MmHg

Pressure in millimeters of mercury at 0°C

Kg/cm2

Pressure in kilograms per square-centimeter

Table 8: Error codes

Error Number

Error Description

100

A non-numeric entry was received where it should be a numeric entry

101

Too many significant digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

106

An invalid pressure unit was received

107

An invalid CJC_STATE was received

108

An invalid TSENS_TYPE was received

109

Pressure module not connected

110

An unknown command was received

111

An invalid RTD or TC parameter value was received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

115

Output is overloaded

116

Calibrator not in pulse train mode when TRIG was received

117

An invalid FREQ_TYPE was received

7.5 Entering Commands

Commands for the calibrator may be entered in upper or lower case.
There is at least one space required between the command and
parameter, all other spaces are optional. Almost all commands for
the calibrator are sequential, any overlapped commands will be
indicated as such. This section will briefly explain each of the
commands and describe their general use, which will include any

39