99

Table 13: Resistance Measurement

Range

Accuracy

(% of reading ± floor)

Ohms low

0.00

W - 400.0W

0.015% ± 0.03

W

Ohms high

401.0

W - 4000.0W

0.015% ± 0.3

W

Table 14: Resistance Source

Range

Excitation Current

Accuracy

(% of reading ± floor)

Ohms low

5.0

W - 400.0W

0.1mA - 0.5mA

0.015% ± 0.1

W

5.0

W - 400.0W

0.5mA - 3mA

0.015% ± 0.03

W

Ohms high

400

W - 1500W

0.05mA - 0.8mA

0.015% ± 0.3

W

1500

W - 4000W

0.05mA - 0.4mA

0.015% ± 0.3

W

Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <=
5ms.

Table 15: Thermocouple Measurement/Source

Range

Accuracy

(% of reading ± floor)

Read (mV)

-10.000mV - 75.000mV

0.02% ± 10

mV

Source (mV)

-10.000mV - 75.000mV

0.02% ± 10

mV

Maximum current output in voltage ranges is 1mA with an output impedance of <= 1

W

Table 16: Thermocouple Read and Source (errors in °C)

TC Type

Range (°C)

Accuracy

J

-210.0 - 0.0

0.4

0.0 - 800.0

0.2

800.0 - 1200.0

0.3

K

-200.0 - 0.0

0.6

0.0 - 1000.0

0.3

1000.0 - 1372.0

0.5

T

-250.0 - 0.0

0.6

0.0 - 400.0

0.2

E

-250.0 - -100.0

0.6

-100.0 - 1000.0

0.2

R

0.0 - 1767.0

1.2

S

0.0 - 1767.0

1.2

B

600.0 - 800.0

1.2

800.0 - 1000.0

1.3

1000.0 - 1820.0

1.5

C

0.0 - 1000.0

0.6

1000.0 - 2316.0

2.3