CALIBRATION
- for Open Channel Flow

1. Before starting the calibration determine:

a)

MAX RANGE = ____________________
(Maximum range = distance from the Sensor to Zero
level point)

b)

MAX LEVEL

= ____________________

(Maximum level of flow through flume or weir)

c)

MIN RANGE

= __________________

(Distance from sensor to Max Level)
Minimum range = MAX RANGE - MAX LEVEL
(must be at least 8" / 20.3 cm depending on sensor
model).

2. Check the maximum range with the sensor installed by:

a)

Use

Æ to get to

UNITS/MODE

then

È to get to

Range

, then

Store? Yes

(NOTE: flow must be at zero flow)
or

b)

Carefully measure distance from sensor to zero level
with tape measure.

AquaMetrix

Ultrason

Page 11

MinRg

1.333 ft

MaxRg

32.00 ft

4mA @

0.00 %

20mA @

100.00 %

CALIBRATION

Damping

20

%

RejTime

08 s

Store?

Yes

RELAY PARAMETERS

FLUME SELECTION

PALMER BOWLUS

SENSOR

ZERO LEVEL

MAX
RANGE

MIN
RANGE

MAX LEVEL

V-NOTCH

SENSOR

ZERO LEVEL

MAX
RANGE

MAX LEVEL

MIN
RANGE