POWER INPUT OPTION
12VDC OR 24VDC

Ultrason Level & Flow Monitors may be ordered factory-configured for 12VDC, or 24VDC power
input.

QUICK BENCH TEST:

Connect Sensor as shown below, then Power. When properly connected a soft clicking can be heard
from the sensor and figures will show on the large LCD display. Test operation of the Ultrason by
holding the sensor steadily and aiming at a flat, stable target 12 to 28" (305 to 711 mm) away from the
end of the sensor. Allow a few seconds for the Ultrason to lock onto the target before displaying its
distance. The Ultrason will now display Range in ft or cm (factory calibration).

CONNECTIONS:

POWER INPUT: Connect only 12VDC/0.5 Amps to the + and - terminals for units marked 12V, or
24VDC/0.5 Amps for units marked 24VDC. The Power Input GND must be connected to the nearest
Ground pole. A 1 amp fuse in line is recommended. Power Consumption is 6.5 W continuous.

AquaMetrix

Ultrason

Page 32

GND

12 AWG

MAX

OPTIONAL 24VDC
OPTIONAL 12VDC

POWER INPUT

IMPORTANT:
MUST CONNECT TO A
GOOD GROUND (

<1 Ohm)

WITH 12 AWG CONDUCTOR

24VDC

12V or

1 AMP FUSE