Run 6, Totalizer (flow mode) 6, Password 6 – Metex 3200U Ultrasonic Flow & Level Monitor User Manual

Page 6: A qua m etrix

Advertising
background image

RUN

A scrolling display shows the units selected from the

UNITS/MODE

column, the mode of operation (

Range, Level

or

Flow

), the full scale value for the large numeric display, Totalizer

value (in

FLOW

mode), and air temperature at the sensor location.

ECHO CONFIDENCE

Echo Strength is automatically adjusted by the Ultrason in response
to operating conditions in the applications.

From RUN use

È to get to the Echo Confidence display,

EC

.

TOTALIZER

The Totalizer display will only be enabled when the Ultrason is calibrated in

FLOW

mode. From

RUN

use

È or Ç keys to display

Tot:

value. The Totalizer value is updated every 2 seconds with flow

volume > 1 litre (0.264 USG). The display will show up to 10 digits and then restart at 0 automatically.

Press

È or Ç keys to return to

RUN

.

The totalizer can be reset to zero by going

Æ to

SPECIAL FUNCTION

and

È to

Reset Tot?

.

RELAY STATUS DISPLAY

Press

È from

Tot:

to

RELAYS 1,2,3

.

Engerized relays will be displayed in reverse type

RELAYS

:

PASSWORD

The password (a number from

00

to

99

) prevents unauthorized access to the

CALIBRATION

menu.

From

RUN

(scrolling display) press

Æ to

PASSWORD

.

Press

Æ to place the cursor under the digits and È or Ç to change the

number and then

Æ to proceed to

CALIBRATION

.

Factory default Password is

00

. A new password can be stored by going

Æ to

SPECIAL FUNCTION

and

È to

New Password

.

AquaMetrix

Ultrason

Page 6

PASSWORD: 00

RUN

FLOW MODE ONLY

RELAYS:1 2 3 4 5 6

Tot:

EC

T: 9999999999

T: 0283168470

TOT:

0

TOT:

0

Units: ft

³, USG, IG, L

Units = m

³

Current

Echo Strength

Arrow

Minimum

Marker

EC

PASSWORD: 00

PASSWORD:

0

0

PASSWORD:

0

1