Field troubleshooting 25, A qua m etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual

Page 25

Advertising
background image

FIELD TROUBLESHOOTING

SYMPTOMS

CHECK

Display - full scale

A

- zero

B

- erratic - random

C

- drifting up

D

- drifting down

E

ECHO LOSS

prompt - flashing

F

Calibration Non-Linear

H

SYMPTOMS

FAULTS

SOLUTIONS

Unit “See’s” Wrong Target Due To:

A,C,D,F

- sensor not aimed correctly

A,F,D

- condensation/dust/dirt buildup on

sensor

- Clean carefully (do not scratch sensor face)

A,D

- sensor mounting stand pipe

- Lower Sensor below stand pipe intrusion

- too long / - too narrow
- dirty / - gasket intruding

C,E

- very turbulent flow in open channel

- Increase

RejTime (CALIBRATION

menu)

- install stilling well on flume or weir

C,E

- very turbulent level in tank

- Increase

RejTime (CALIBRATION

menu)

- change tank fill method

Unit Picks-Up Interference Due To:

A,C

- noise from high pressure fill

- Install submerged fill pipe

A,D

- sensor coupling over tightened

- Hand tighten only (like a light bulb)

A,D

- sensor coupling not used

- Use coupling supplied

C

- other ultrasonic unit in close proximity

- Synchronize

Electrical interference:

C

- sensor cable connections reversed

C,G

- through sensor cable

- Use properly grounded metal conduit

C

- sensor cable extended and junction not

insulated

- Use metal Junction Box

C,G

- through enclosure

- Use metal enclosure

C,G

- through 4-20mA output cable

- Use shielded twisted pair (shielded to AC

ground)

- Use grounded metal conduit

C,G

- wiring or installation close to variable

speed drive or inverter

- Follow V.S.D. manufacturer’s instructions

for Drive grounding, wiring and shielding

AquaMetrix

Ultrason

Page 25