FLUME SELECTION

The

FLUME SELECTION

menu will only appear in

FLOW

mode.

Use

È and Æ to select the correct Flume or Weir.

Use

È and Æ again to select the Size then È to

Store?

and

Æ to

Yes

.

Press

È or Ç to store flume selection and get to

CALIBRATION

mode.

CUSTOM FLUMES: Experimental or empirical data for any flume or weir can be reduced to an
equation of the form

Q = K Hn

where:

Q

is flow volume

K

is as scaling factor which includes flume size and units of flow measurement

n

describes the flume non-linearity

H

is the level being measured

Use the AquaMetrix Utility program

FIND K&n.exe

to determine

K

and

n

constants from entered data.

Select

CUSTOM

for flumes or weirs not listed in the Menu. At the

K:

prompt enter a 'K' constant. At the

n:

enter an 'n' constant.

AquaMetrix

Ultrason

Page 9

Store?

Yes

CALIBRATION

FLUME SELECTION

FLOW MODE ONLY

V-Notch

More>

P. Bowlus More>

L.Lagco

More>

Rect. Weir

>

Rect.Weir w/EC >

Custom

More>

22.5 30 45 60

0

9

4"

6"

8"

10">

12"

15"

18">

21"

24"

42">

6" 8" 10" 12" >

18" 24" 30" 36">

Size 1.000 ft

Size 1.000 ft

K =

1.0000

N =

1.0000

Parshall

More>

2"

3"

6"

9" >

12" 18" 24" 36">

UNITS/MODE