NOSHOK 2100 Series Field Upgradeable Dual Input Process Indicator User Manual

Page 4

Advertising
background image

4

16. CERTIFICATIONS AND COMPLIANCES:

SAFETY

Type 4X Enclosure rating (Face only), UL50
IEC 1010-1, EN 61010-1: Safety requirements for electrical equipment for

measurement, control, and laboratory use, Part I

IP65 Enclosure rating (Face only), IEC 529
IP20 Enclosure rating (Rear of unit), IEC 529

ELECTROMAGNETIC COMPATIBILITY

Emissions and Immunity to EN 61326: Electrical Equipment for

Measurement, Control and Laboratory use.

Notes:

1. Criterion A: Normal operation within specified limits.
2. Criterion B: Temporary loss of performance from which the unit self-
recovers.

Refer to EMC Installation Guidelines section of the bulletin for additional

information.

17. CONNECTIONS: High compression cage-clamp terminal block

Wire Strip Length: 0.3" (7.5 mm)
Wire Gage: 30-14 AWG copper wire
Torque: 4.5 inch-lbs (0.51 N-m) max.

18. CONSTRUCTION: This unit is rated for NEMA 4X/IP65 outdoor use.

IP20 Touch safe. Installation Category II, Pollution Degree 2. One piece
bezel/case. Flame resistant. Synthetic rubber keypad. Panel gasket and
mounting clip included.

19. WEIGHT: 10.4 oz. (295 g)

Immunity to Industrial Locations:
Electrostatic discharge

EN 61000-4-2

Criterion A
4 kV contact discharge
8 kV air discharge

Electromagnetic RF fields

EN 61000-4-3

Criterion A
10 V/m

2 kV L&N-E power

Fast transients (burst)

EN 61000-4-4

Criterion A
2 kV power
1 kV signal

Surge

EN 61000-4-5

Criterion A
1 kV L-L,

DC powered

RF conducted interference

EN 61000-4-6

Criterion A
3 V/rms

EN 55011

Class A

Emissions:
AC powered

EN 55011

Class B

Advertising