background image

C.3

TRICOR Systems Inc.

Figure C.1 921A Rear Panel

Press TEST/READY again -

The platform returns to the home

position. The resistance Offset
Adjustment Mode message is
displayed.

Install the Offset/Gain Jumpering Connector (part of PN 921-
811-XX) on the Resistance Cable (PN 921-112) P2 Connector.
Do not remove the jumpering connector until instructed to
do so by this procedure.

Step 5

Press the TEST/READY switch once. The unit displays the
offset values for the resistance current source channel. Figure
C.2 identifies the location of the offset potentiometer (pot) and
the corresponding pot designator which is displayed by the unit.

Step 6

The resistance offset value must be within the acceptable range
listed below. If necessary, adjust the offset pot until the offset
value displayed by the unit is within the acceptable range.

Rear Panel