SHIMPO 921A User Manual

Page 46

Advertising
background image

C.6

TRICOR Systems Inc.

Table C.2 4

5, 10 mA or 25 mA Resistance Option using Resistance

Fixture PN 921-811-09

Resistance Switch Position

Displayed Value (

5)

Tolerance

1 (0

5)

0.000

±.010

2 (1

5)

1.000

±.010

3 (2

5)

2.000

±.010

4 (3

5)

3.000

±.010

5 (3.8

5)

3.800

±.010

6 (.025

5)

0.025

±.010

7 (.050

5)

0.050

±.010

8 (.075

5)

0.075

±.010

Table C.3 16 k

5, .25 mA Resistance Option using Resistance Fixture

PN 921-811-01

Resistance Switch Position

Displayed Value*

Tolerance

0 (0 k

5)

0.000

± 0.010

1 (4 k

5)

1.000

± 0.010

2 (8 k

5)

2.000

± 0.010

3 (12 k

5)

3.000

± 0.010

4 (16 k

5)

4.000

± 0.010

* Resistance = Displayed Value x 4000 (

5)

Table C.4 1 k

5, 1 mA Resistance Option using Resistance Fixture

PN 921-811-07

Resistance Switch Position

Displayed Value*

Tolerance

0 (0

5)

0.000

± 0.010

1 (250

5)

1.000

± 0.010

2 (500

5)

2.000

± 0.010

3 (750

5)

3.000

± 0.010

4 (1000

5)

4.000

± 0.010

* Resistance = Displayed Value x 250 (

5)