P7.3 Selecting Automatic or Manual Temperature Compensation
This feature lets you select between Automatic
Temperature Compensation (ATC) and Manual
Temperature Compensation. Meter default is ATC.
From measurement mode
1. Press the MODE key to select conductivity
measurement mode.
2. Press
SETUP key to enter Set Up mode.
3. Press
the or
keys to scroll through subgroups
until you view parameter P7.0.
4. Press
the
ENTER key three times to select
parameter 7.3. The upper display shows “ATC” and
the lower display shows “P7.3”.
5. Press
the
ENTER key again. The upper display
shows “ATC” and the lower display shows “YES” or
“NO”.
6. Press
the or
keys to select the Automatic
Temperature Compensation on or off.
•
YES = ATC on; NO = ATC off
7. Press
the
ENTER key to confirm selection and
proceed to P7.4 next page. Press the CAL/MEAS
key to return to measurement mode.
Figure 38 - Selecting ATC on or off