Thermo Fisher Scientific CyberScan PC 300 User Manual
Page 53

Instruction Manual PC 300
49
7.9.1
P9.0: Resetting to factory default settings (conductivity)
Program 9.0 lets you reset all parameters to factory
default settings. This clears all calibration data and
any other conductivity setup functions you might
have changed.
From measurement mode
1. Press
the
MODE key to select conductivity
measurement mode.
2. Press
SETUP key to enter Set Up mode.
3. Press
the or
keys to scroll through
subgroups until you view parameter P9.0.
4. Press
the
ENTER key to enter parameter P9.0.
5. Press
the or
keys to toggle between NO
and YES.
•
NO retains current settings; YES resets
to factory default settings.
6. Press
the
ENTER key to confirm selection and
to return to the measurement mode. Otherwise
press
CAL/MEAS key to return to
measurement mode without resetting to factory
default.
See Figure 42.
Notes:
To clear all pH data, see Section 7.5.
See Addendum 4 for a table of factory default settings.
Figure 42 – P9.0: Reset to factory
default values