Instruction Manual PC 300
48
P8.2 Adjusting the normalization temperature
Your meter will normalize its conductivity
measurements to a standard temperature that you can
select. You can adjust the normalization temperature
from 15 to 30
°C (59 to 86 °F). Meter default is 25 °C
(77
°F).
From measurement mode
1. Press
the
MODE key to select conductivity
measurement mode.
2. Press
SETUP key to enter Set Up mode.
3. Press
the or
keys to scroll through
subgroups until you view parameter P8.0.
4. Press
the
ENTER key three times to select
parameter 8.2. The display shows “t.nr” on the
upper display.
5. Press
the
ENTER key again. The upper display
shows the normalization temperature and the
lower display shows “t.nr”.
6. Press
the or
keys to select the
normalization temperature.
7. Press
the
ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS
key to return to measurement mode.
Figure 41 - Adjusting the
normalization temperature