Thermo Fisher Scientific CyberScan pH/Ion 510 User Manual

Page 23

Advertising
background image

Instruction Manual

pH 510/ Ion 510

-

19

-

5.3

P3.0 Selecting pH Buffer Standard and No of Calibration Points



The available standards are USA and NIST.

1. Enter the SETUP mode as described in this section earlier. The

meter automatically goes to program P1.0.

2. Press

¿ or À key to scroll through the programs until you view “bUF

in the upper display and P3.0 in the lower display. The buffer
annunciator also appears.

3. Press

ENTER key to access the program P3.0. The lower display will

read either ”USA” or “nSt (depending on the previous setting you
have made; factory default is USA.)

4. Press

¿ or À key to select the buffer set you require:

USA: 1.68, 4.01, 7.00, 10.01, 12.45

NIST:

1.68, 4.01, 6.86, 9.18, 12.45

5. To confirm the buffer standard, press ENTER. You will now enter the

selection mode for calibration points.

6. Press

¿ or À key to select the calibration points that you desire. You

can now select from 2 to 5 calibration points.

7. Press

ENTER to confirm your selection.

8. To exit P3.0 program press

¿ or À key to select a new program OR

press the CAL/MEAS to return to the measurement mode.