7 troubleshooting guide – Thermo Fisher Scientific CyberScan pH/Ion 510 User Manual

Page 28

Advertising
background image

Instruction Manual

pH 510/ Ion 510

-

24

-

7 TROUBLESHOOTING GUIDE

7.1 Error

Messages

The table provides a quick guideline for diagnosis of possible problems indicated by the messages generated by the 510
meter series. The table also provides possible solutions to problems encountered.
Diagrams inserted below.

Display Screen

ERROR

MESSAGE

INDICATES CAUSE CORRECTIVE

ACTION

Err. 2 (in upper
display) during Ion
calibration

Too few
calibration points

Meter calibrated with
only 1-point in ion
calibration

Recalibrate using 2 or
more points

Err. 3 (in upper
display) during Ion
calibration

Slope error

Ion calibration
solution not within
acceptable range
(15mV to
90mV/decade)

Recalibrate using fresh
ion solutions

Replace electrode.

Err. 4 (in upper
display) during Ion
calibration

Calibration points
more than 1
decade apart

Calibration points
more than 1 decade
apart

Recalibrate using
calibration points 1
decade apart.

Err (annunciator) Wrong

keypad

input

Button does not work
in the current
operation mode

Release key. Select
valid buttons
depending on mode.

Electrode icon
(indicator)

Calibration error

Buffer value does not
match value displayed
or electrode is
disconnected or
failing

Use fresh buffer
solution or check
electrode connection.

Possibly replace
electrodes.

Ur or Or (in upper
display)

Measured value is
out of range

Electrode not
connected

Electrode clogged,
dirty or broken

Meter not calibrated.

Measurement is
under range or over
range

Make sure electrode is
connected

Clean or replace
electrode

Recalibrate the meter.

Treat samples to bring
within meter measuring
range