Thermo Fisher Scientific CyberScan pH/Ion 510 User Manual
Page 24

Instruction Manual
pH 510/ Ion 510
-
20
-
5.4
P4.0 Meter Reset
This program gives you the option to either reset the calibration data’s of
all parameters or reset everything back to factory default condition except
for temperature calibrated value and buffer set selection (NIST or USA).
1.
Enter the SETUP mode as described in this section earlier on. The
meter automatically goes to program P1.0.
2. Press
¿ or À key to scroll through the programs until you view rSt in
the upper display and P4.0 in the lower display.
3. Press
the
ENTER key to access the program P4.0. The lower
display will read NO.
4. Press
¿ or À key to select “NO” or “YES”
• NO - deactivates the reset function
• YES – Activates the reset option mode
5. Press
ENTER to confirm your selection.
6.
For reset option mode, press
¿ or À key to select “CAL” or “FCt”
• CAL – clears all calibration datas in all parameters
• FCt – clears everything back to factory default condition except
for temperature and buffer set selection (NIST or USA)
7. Press
ENTER to confirm your reset option. The lower display will
flash for a few seconds before the meter returns to program P4.0.
SETUP
SETUP
SETUP
YES
SETUP
CAL
ENTER
ENTER