Thermo Fisher Scientific CyberScan pH/Ion 510 User Manual

Page 24

Advertising
background image

Instruction Manual

pH 510/ Ion 510

-

20

-



5.4

P4.0 Meter Reset




This program gives you the option to either reset the calibration data’s of
all parameters or reset everything back to factory default condition except
for temperature calibrated value and buffer set selection (NIST or USA).

1.

Enter the SETUP mode as described in this section earlier on. The
meter automatically goes to program P1.0.

2. Press

¿ or À key to scroll through the programs until you view rSt in

the upper display and P4.0 in the lower display.

3. Press

the

ENTER key to access the program P4.0. The lower

display will read NO.

4. Press

¿ or À key to select “NO” or “YES

NO - deactivates the reset function

YES – Activates the reset option mode

5. Press

ENTER to confirm your selection.

6.

For reset option mode, press

¿ or À key to select “CAL” or “FCt

CAL – clears all calibration datas in all parameters

FCt – clears everything back to factory default condition except

for temperature and buffer set selection (NIST or USA)

7. Press

ENTER to confirm your reset option. The lower display will

flash for a few seconds before the meter returns to program P4.0.

SETUP

SETUP

SETUP

YES

SETUP

CAL

ENTER

ENTER