7.2 Troubleshooting

PROBLEM PROBABLE

CAUSE

REMEDIAL

ACTION

No display

a) Mains power not switched on.

b) AC Adapter socket not inserted properly.

a) Switch on the power supply.

b) Re-insert AC Adapter socket.

Unstable reading

a) Insufficient reference electrolyte in

electrode.

a) Fill electrode with reference electrolyte.

b) Broken electrode

b) Replace electrode.

c) External ‘noises’ or induction ( e.g.

electrical ‘noise’ caused by a nearby
motor)

c) Remove or switch off interfering device.

d) Dirty electrode.

d) Clean the electrode. Hydrate, then rejuvenate if

necessary.

Slow response

a) Dirty electrode.

a) Clean electrode. Hydrate, then rejuvenate if

necessary.

Meter not
responding to key
press

a) HOLD mode in operation.

b) Internal program error.

a) Cancel HOLD mode. Press Hold button.

b) Reset all internal programs by removing and

reinserting the A/C adapter.