Thermo Fisher Scientific CyberScan pH/Ion 510 User Manual
Page 33

Instruction Manual
pH 510/ Ion 510
-
29
-
8.5 Specifications
Description
pH 510
Ion 510
pH Range
0.00 to 14.00 pH
9
9
Resolution
0.01 pH
9
9
Accuracy
± 0.01 pH + 1 count
9
9
Ion Concentration Range
0.01 to 1999 ppm
9
(± 500mV input)
Resolution
0.01 pm (0.01 to 0.99 ppm); 0.1 ppm (1.0
to 199.9 ppm); 1 ppm (200 to 1999 ppm)
9
Accuracy
±0.5% Full Scale for mono-valent ions
±1.0 % Full Scale for di-valent ions
9
mV Range
Yes
9
± 1999 mV
9
± 500mV(Absolute)
±1999mV( Relative)
Resolution
0.1 mV (
± 199.9 mV), 1 mV beyond ± 200 mV
9
9
Accuracy
± 0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200
mV
9
9
Calibration Offset
Adjustment
Up to ±150 mV
9
9
Temperature Range
0.0 to 100.0
o
C
9
9
Resolution
0.1
o
C
9
9
Accuracy
± 0.3
o
C
9
9
Temperature Compensation
Automatic / Manual (0 to 100
o
C)
9
9
No. of pH Calibration Points
Up to 5 points with Auto-buffer recognition
9
9
pH Buffer Calibration Options
USA : pH1.68, 4.01, 7.00, 10.01,12.45
NIST: pH1.68, 4.01, 6.86, 9.18 ,12.45
9
9
No. of Ion Calibration Points
Up to 3 points (minimum 2 points)
9
Ion Calibration Options
0.1, 1.0, 10.0, 100.0 or 1000.0 ppm
9
Min. & Max. Slope during
Calibration
15 mV/decade to 90 mV/decade
9
Temperature Calibration
Offset in 0.1
o
C increments; Offset range:
± 5
o
C
9
9
Memory
100 data sets
9
9
Hold Function
Yes
9
9
Self-diagnostic Messages
Yes
9
9
pH Slope and Offset Display
Yes
9
9
Inputs
BNC, Phono, Ref. (Half-cell), Ground
9
9
Output
Recorder (+/- 2000 mV)
9
9
Power Requirements
AC/DC 9V Adapter (110 VAC/220 VAC, 50-
60Hz)
9
9
Display
Custom Dual LCD (1 x 4 digits, 1 x 3.5 digits,
annunicators)
9
9
Dimensions / Weight
Meter with electrode holder: 18 x 23 x 6 cm /
1250 g
Boxed (meter with electrode holder): 40 x 26 x 9
cm / 1950 g
9
9