Thermo Fisher Scientific CyberScan pH/Ion 510 User Manual

Page 33

Advertising
background image

Instruction Manual

pH 510/ Ion 510

-

29

-

8.5 Specifications

Description

pH 510

Ion 510

pH Range

0.00 to 14.00 pH

9

9

Resolution

0.01 pH

9

9

Accuracy

± 0.01 pH + 1 count

9

9

Ion Concentration Range

0.01 to 1999 ppm

9

(± 500mV input)

Resolution

0.01 pm (0.01 to 0.99 ppm); 0.1 ppm (1.0

to 199.9 ppm); 1 ppm (200 to 1999 ppm)

9

Accuracy

±0.5% Full Scale for mono-valent ions

±1.0 % Full Scale for di-valent ions

9

mV Range

Yes

9

± 1999 mV

9

± 500mV(Absolute)

±1999mV( Relative)

Resolution

0.1 mV (

± 199.9 mV), 1 mV beyond ± 200 mV

9

9

Accuracy

± 0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200

mV

9

9

Calibration Offset
Adjustment

Up to ±150 mV

9

9

Temperature Range

0.0 to 100.0

o

C

9

9

Resolution

0.1

o

C

9

9

Accuracy

± 0.3

o

C

9

9

Temperature Compensation

Automatic / Manual (0 to 100

o

C)

9

9

No. of pH Calibration Points

Up to 5 points with Auto-buffer recognition

9

9

pH Buffer Calibration Options

USA : pH1.68, 4.01, 7.00, 10.01,12.45

NIST: pH1.68, 4.01, 6.86, 9.18 ,12.45

9

9

No. of Ion Calibration Points

Up to 3 points (minimum 2 points)

9

Ion Calibration Options

0.1, 1.0, 10.0, 100.0 or 1000.0 ppm

9

Min. & Max. Slope during
Calibration

15 mV/decade to 90 mV/decade

9

Temperature Calibration

Offset in 0.1

o

C increments; Offset range:

± 5

o

C

9

9

Memory

100 data sets

9

9

Hold Function

Yes

9

9

Self-diagnostic Messages

Yes

9

9

pH Slope and Offset Display

Yes

9

9

Inputs

BNC, Phono, Ref. (Half-cell), Ground

9

9

Output

Recorder (+/- 2000 mV)

9

9

Power Requirements

AC/DC 9V Adapter (110 VAC/220 VAC, 50-

60Hz)

9

9

Display

Custom Dual LCD (1 x 4 digits, 1 x 3.5 digits,

annunicators)

9

9

Dimensions / Weight

Meter with electrode holder: 18 x 23 x 6 cm /

1250 g

Boxed (meter with electrode holder): 40 x 26 x 9

cm / 1950 g

9

9

Advertising