20 configuration 3.1, Mode display code function description – West Control Solutions MRC 8000 User Manual

Page 20

Advertising
background image

20

Configuration 3.1

After completing installation and wiring of the instrument the configuration (set up) procedures
must be performed to prepare the instrument for operation on the intended application. The
procedures include selecting specific parameters, entering data and possible jumper position-
ing. Once properly configured the instrument will retain the user selections in memory so this
procedure need not be repeated unless required by changes in the application.
Parameter selections and data entry are made via the front keypad. To ease configuration
and operation, user entered data has been divided up into several sections referred to as
modes. Each mode contains a different type of data or may be used for specific operating
functions. For two pen instruments, some modes are common to both pens. These modes are
as follows:

RECORDER-

RECORDING CONTROLLER-

MODE

DISPLAY CODE

FUNCTION

DESCRIPTION

Off

oFF

Operation

Outputs and Alarms
are Off
Chart may stop
rotating(selectable)

Operate* oPEr Operation

Limits and Alarms
are Active

Control**

CtrL

Control

Outputs and Alarms
are Active

Test

tESt

Service

Tests Instrument
Operation

Calibration

CAL

Service

Calibrates, Resets
Instrument

Program

Prog

Configuration

Configure Operating
Parameters

Alarm Set* ASEt Configuration &

Enter Alarm Settings

Operation

Tune**

tunE

Configuration &

Enter Tune and

Operation

Alarm Settings

Setpoint Selection**

SPS

Operation

Selects Remote or
Local Setpoint
Operation (Remote
Setpoint Optional)

* Applies to Recorders
** Applies to Recording Controllers

Operation
(oPEr)

Test
(tESt)

Program
(Prog)

Alarm Set
(ASEt)

Calibrate
(CAL)

Control
(CtrL)

Test
(tESt)

Program
(Prog)

Tune
(tunE)

Setpoint Select
(SPS)

Calibrate
(CAL)

Standby
(Stby)

OFF
(oFF)

Advertising