Troubleshooting – Badger Meter Cox Precision Turbine Meters User Manual

Page 10

Advertising
background image

TROUBLESHOOTING

Issue

Possible Cause

Remedy

Meter indicates higher flow than actual . Cavitation .

Increase back pressure .

Meter indicates high flow .

Dirt blocking flow area rotor .

Clean meter; add filter .

Meter indicates low flow .

Dirt dragging rotor .

Clean meter; add filter .

Meter indicates low flow .

Worn bearing .

Replace bearing; recalibrate when

required .

Meter indicates low flow .

Viscosity higher than calibrated .

Change temperature; change fluid;

recalibrate meter .

Erratic system indication; meter alone

works well .

Ground loop is shielding .

Ground shield one place only . Watch for

internal electronic instrument grounds .

Indicator shows flow when shut off .

Mechanical vibration causes rotor to

oscillate without turning .

Mechanical vibration .

Isolate meter; use potted pickoff .

No flow indication . Full flow of fluid

opened into dry meter . Impact of fluid

on rotor causes bearing separation .

Fluid shock . New bearing failed .

Move meter to position where it is full

of fluid at start-up .

Erratic indication at low flow; good

indication at high flow .

Low instrument sensitivity . 10

mV rms turbine signal is being

lowered by loading of electronics or

instrumentation cannot sense low level

signals .

Amplify signal .

No flow indication .

Faulty pickoff .

Replace pickoff; recalibrate as

necessary .

System works perfectly, except

indicates lower flow over entire range .

Bypass flow, leak .

Eliminate bypass valves, leak . Faulty

solenoid valves .

Meter indicating high flow . Upstream

piping at meter smaller than meter .

Fluid jet impingement on rotor . Critical

in gas .

Change piping .

Opposite effects as above .

Viscosity lower than calibrated .

Change temperature; change fluid;

recalibrate meter .

Mass flow indication wrong . Turbine

meter is volumetric; density correction

is electronic; must change with

temperature .

Wrong fluid density . Critical in gas .

Check fluid, electronics .

Erratic or wrong indication of flow .

Loose pickoff .

Tighten pickoff .

Indicates high flow two hours after

installing new bearing .

Bearing wear-in; small meters critical .

Recalibrate . 20…30 min . run-in is

required to stabilize friction .

Cannot reach maximum flow rate;

meter selection was with Delta-P at 0 .75

sp . gr ., now using on 1 .0 sp . gr . Delta-P

is proportional to specific gravity .

High pressure drop .

Install larger meter .

Does not repeat at low flows . Repeats at

high flows .

System resolution readability .

Increase resolution, for example:

1 out of 100 = 1%

1 out of 1000 = 0 .1%

Table 1: Troubleshooting