RECALIBRATION

• Recalibration is not necessary following a cleaning operation or the replacement of bearings, snap rings, springs

or spacers .

• Recalibrate the flow meter if the rotor hub, or rotor and flow straightener assembly is replaced .
• Flow meters may be recalibrated by the user if the facilities are available, or they may be returned to the factory . Yearly

calibration is recommended .

• When the flow meter is set up for recalibration, allow the fluid to circulate for 5 minutes before beginning the

calibrating runs .

LOFLO REPLACEMENT PARTS

Figure 20: Model CLFA6

Figure 21: Model CLFB6

Figure 22: Models CLFC6…CLFF6

1100 – Flow is through 0 .03
1300 – Rotor is angled blade
1400 – Bearing is staked into rotor .

(Factory replace)

1100 – Flow is through swirl slot .

Tube is press fit over slot .
1300 – Rotor has straight blades .
1400 – Bearing is staked into rotor .

(Factory replace)

1100 – Flow is through swirl slots . No tube over slots .
1300 – Rotor has straight blades .
1400 – Bearing is held with snap rings . (Field replace)