Procedure for Cleaning a Turbine Meter after Water Calibration and/or Service

OTEE:

N

When cleaning flow meters, keep the body, sleeve and pickoff together . Sleeve is fitted to body and pickoff has a

protruding pin . Replacement pickoffs are supplied with a nut and have no protruding pin .

1 . Remove the meter from the line and let all excess water drip out .
2 . Fill the meter with alcohol (at least 50% Isopropyl, Ethyl or Methyl) and let it stand for 5 minutes .
3 . Discard the alcohol and let the meter dry for 2 minutes .
4 . Fill the meter with MIL-C-7024 Type 2 calibration fluid (or similar solvent) and let it stand for 1 minute .
5 . Discard the calibration fluid and flush the meter with an approved fluorocarbon solvent, such as Isotron .

OTEE:

N

If this procedure is not possible, the turbine meter should always remain filled with water when not in use, to prevent

internal wetted parts from being exposed to air .

CLFC6, CLFD6, CLFE6

CLFB6, CLFA6

CLFF6

CLFC6, CLFD6, CLFE6, CLFF6
1 . Body *

2 . C-washer

3 . Hub *

4 . Spring *

5 . Spacer

6 . Bearing retainer ring *

7 . Bearing

8 . Rotor *

9 . Straightener assembly *

10 . Body retaining ring

CLFB6, CLFA6
1 . Body *

2 . C-washer

3 . Hub *

4 . Spring *

5 . Spacer

8 . Rotor *

9 . Straightener assembly *

10 . Body retaining ring

* Replaceable at factory level only .

OTEE:

N

Order parts by name and basic number shown, followed by flow meter

size designation . For example, order a bearing for a CLFF6 flow meter

as bearing #1400-63 . Flow meter serial number must be provided when

ordering parts .

Figure 2: Cox Precision LoFlo turbine flow meter

Figure 3: Cox precision turbine flow meter

DO NOT INTERCHANGE FLOW METER PARTS OTHER THAN BEARINGS AND RETAINING RINGS. THIS PRECAUTION IS

NECESSARY TO PRESERVE LINEARITY AND REPEATABILITY.