Spectrophotometer unit specifications, Specifications, Specifications 8 – Canon imagePROGRAF iPF6450 User Manual
Page 1038: Spectrophotometer unit specifications 8

Spectrophotometer Unit Specifications
Spectrophotometer Unit Specifications
Specifications
Specifications
Important
•
The following values may vary depending on the operating environment.
Spectrophotometer Sensor
Model name
X-Rite ILS25
Measurement geometry conditions
0°/45° ANSI/ISO 5-4
Illumination
Tungsten lamp with UV cut filter + UV-LED
Measurement aperture
Diameter 3.55 mm
Spectral Output
31 points (400nm to 700nm 10nm gap)
Observer Angle
2°/10°
Illumination Type
A, C, D50, D55, D65, D75, F2, F7, F8, F10, F11, F12
Measurement Range
0 to 150% reflection
Repetition Accuracy
Maximum dEcmc
≦0.2 (White tile)
Maximum dEcmc
≦0.6 (Black tile)
Instrument Error
Average dEcmc
≦0.5 (BCRA tile 12 color)
Max dEcmc ≤ 1.0 (12-color BCRA tiles)
Recommended calibration frequency
Once per year
Specifications
iPF6450
Specifications
Spectrophotometer Unit
Specifications
1038