Agilent Technologies 664xA User Manual

Page 31

Advertising
background image

Language Dictionary 31

*RST


Meaning and Type
Reset Device State

Description
This command resets the power supply to a factory-defined state as defined below. *RST also forces an ABORt command.

Command State

CAL:STAT OFF

OUTP[:STAT] OFF

CURR[:LEV][:IMM] *

OUTP:PROT:DEL *

CURR[:LEV]:TRIG *

OUTP:REL[:STAT] OFF

CURR:PROT:STAT OFF

OUTP:REL:POL NORM

DIG:DATA 0

TRIG:SOUR BUS

DISP[:WIND]:STAT ON

VOLT[:LEV][:IMM] *

DISP[:WIND]:MODE NORM

VOLT[:LEV][:TRIG] *

DISP[:WIND]:TEXT

VOLT:PROT[:LEV] *

INIT:CONT OFF

* Model-dependent value. See Table 3-1.


Command Syntax

*RST

Parameters

(None)

Query Syntax

(None)

Related Commands

*PSC

*SAV

*SAV


Meaning and Type
SAVE Device State

Description
This command stores the present state of the power supply to the specified location in memory. Up to four states can be
stored. Under certain conditions (see "Turn-On Conditions" in "Chapter 5 - Front Panel Operation" of the Operating
Guide), location 0 may hold the device state that is automatically recalled at power turn-on.
The following power supply parameters are stored by *SAV:

CURR[:LEV][:IMM]

OUTP[:STAT]

OUTP:REL:POL

CURR:PROT:STAT

OUTP:PROT:DEL

VOLT[:LEV][:IMM]

DIG:DATA[:VAL]

OUTP:REL[:STAT]

VOLT:PROT[:LEV]


Command Syntax

*SAV <NRf>

Parameters

0|1|2|3

Example

SAV 3

Query Syntax

(None)

Related Commands

*RCL *RST


The power supply uses nonvolatile memory for recording register states. Programs that repeatedly use
*SAV for recalling states cause frequent write cycles to the memory and can eventually exceed the
maximum number of write cycles for the memory (see in the power supply Operating Guide).

Advertising
This manual is related to the following products: