40 Language Dictionary

OUTP:REL

Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.


This command is valid only if the power supply is configured for the optional relay connector. Programming ON closes the
relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the power
supply is supplying power to a load, that power will appear at the relay contacts during switching. If the power supply is not
configured for the relay option, sending either relay command generates an error.

Command Syntax

OUTPut:RELay[:STATe] <bool>

Parameters

0 | 1 | OFF | ON

*RST Value

0

Examples

OUTP:REL 1 OUTP:REL OFF

Query Syntax

OUTPput:RELay?

Returned Parameters

0 |1

Related Commands

OUTP[:STAT] *RCL *SAV

OUTP:REL:POL

Do not install or program the Agilent Relay Accessories if the power supply maximum output current
rating (see Table 3-1) exceeds the contact ratings of the relay.


This command is valid only if the power supply is configured for the optional relay connector. Programming NORMal
causes the relay output polarity to be the same as the power supply output. Programming REVerse causes the relay output
polarity to be opposite to that of the power supply output. If OUTP[:STAT] = ON when either relay command is sent, the
power supply output voltage is set to 0 during the time that the relays are changing polarity.

If the power supply is not configured for the relay option, sending either relay command generates an error.

Command Syntax

OUTPut:RELay:POLarity <CRD>

Parameters

NORMal | REVerse

*RST Value

NORM

Examples

OUTP:REL:POL NORM

Query Syntax

OUTPput:RELay:POLarity?

Returned Parameters

NORM | REV

Related Commands

OUTP[:STAT] *RCL *SAV