Description of subsystem commands, Abor, Calibration commands – Agilent Technologies 664xA User Manual

Page 34

Advertising
background image

34 Language Dictionary

Description of Subsystem Commands

Figure 3-2 is a tree diagram of the subsystem commands. Commands followed by a question mark (?) take only the query
form. Except as noted in the syntax descriptions, all other commands take both the command and query form. The
commands are listed in alphabetical order and the commands within each subsystem are grouped alphabetically under the
subsystem.

Figure 3-2. Subsystem Commands Tree Diagram

ABOR


This command cancels any trigger actions presently in process. Pending trigger levels are reset equal to their corresponding
immediate values. ABOR also resets the WTG bit in the Operation Condition Status register (see "Chapter 4 - Status
Reporting"). If INIT:CONT ON has been programmed, the trigger subsystem initiates itself immediately after ABORt,
thereby setting WTG. ABOR is executed at power turn on and upon execution of *RCL or RST.

Command Syntax

ABORt

Parameters

(None)

Examples

ABOR

Query Syntax

(None)

Related Commands

INIT *RST *TRG TRIG

Calibration Commands


See Appendix A in the power supply Operating Guide.