DEWPOINT MEASUREMENT

The following recommendations should be taken into account when
measuring in very dry environments.

- A clean environment is always best for humidity measurements.

- The number of connections should be minimized to avoid leaks.

- The flow rate must be adequate.

- Dead ends must be avoided as they cannot be flushed easily.

- The tube temperature must never lie under the dewpoint of the

sample gas. This may lead to condensation and false results.

- The sample tubing should be of as short length as possible. The

surface area should be minimized using the narrowest tubing that
the flow conditions allow.

- The surface finish of the pipework is important. Polished or

electro-polished steel is recommended for best results.

- Hygroscopic materials should be avoided in the sampling lines. Use

stainless steel membranes instead of rubber membranes.

- Impermeable materials should be selected to avoid inward diffusion

of moisture through the sampling tubes and enclosures. Such
impermeable materials are, for example, high-quality stainless steel
and metals. Avoid PVC or nylon tubes.

Measuring

If you start DM70 for the first time, see section Preparations Before
Use on page 13
. Otherwise, follow the instructions below.

1.

Connect the probe to the MI70 indicator.

2. Press

the

Power On/Off button.

VAISALA _______________________________________________________________________ 17