Turning off Automatic Auto-Calibration

NOTE

If the automatic auto-calibration is turned off, the auto-calibration
should be started when starting measuring after the probe has not
been used for a while and always at least once every hour.

The automatic auto-calibration can be turned off as follows:

1.

Open the menu by pressing

Open.

2. Select

Settings and press

.

3. Select

Measurement settings and press

.

4.

To turn off the automatic auto-calibration, press

Off. To reactivate the automatic auto-calibration, press

On.

5.

To return to the basic display, press

Exit.

Manual Auto-Calibration

To check that auto-calibration has taken place for obtaining the most
accurate measurement in a very dry environment, you can auto-
calibrate manually before measuring as follows:

1.

Open the menu by pressing

Open.

2. Select

Functions and press

.

3. Select

AutoCal and press

Start.

4.

If you want to start auto-calibration, press Yes. If the auto-
calibration conditions are not fulfilled, a note appears on the
display informing that the calibration cannot be made or that a
new attempt will be made later.

5.

When the auto-calibration is completed, the display returns to
the basic display automatically.

Auto-calibration takes about 2 minutes. When the auto-calibration is
in progress, an estimate of the measurement readings may appear on
the display. Messages concerning the remaining auto-calibration time
are shown periodically. See Figure 9 on page 29.

NOTE

If it is too humid, auto-calibration will not start.

28 __________________________________________________________________ M010091EN-F